The 6340 ComboSource offers excellent low current performance, with both a 4 Amp and 2 Amp operating range, works well for high power pump modules and other medium current diode lasers. With 1mA set point resolution and less than 30uA of noise, the 6340 provides a transparent driver that won’t interfere with the performance of your laser or LED devices.
Dual Range
One of the valuable features of the 6340 is the dual range capability, which allows operation in either a 2A range or 4A range. For applications that need less than 2A, the lower range offers higher accuracy, lower noise, and improved stability, while still maintaining the 4A range for higher power devices.
Ground Loops Eliminated
The 6340 optically and electrically isolates it’s isolated inputs and outputs, preventing device-damaging ground loops. No other driver on the market has this capability.
User Function Keys
The user function keys can be used to quickly select a different configuration state or execute a predefined set of commands. Switch between two different experiments or script repetitive actions…anything you can do manually with the instrument can be programmed to the function key.
Powerful 60 Watt Temperature Controller
In addition to the high-performance diode driver, the 6340 adds a 60 watt temperature controller, giving you substantial power for demanding applications. Only need a little bit of power? No problem, because the 6340 works just as well driving 2 watts of power as it does driving 50. Support for all standard sensors (thermistor, RTD, AD590, and LM335) is also included.
AutoTune Automatic PID Calculation
All Arroyo Instruments temperature controllers feature AutoTune for automatic PID calculation. With AutoTune, you no longer need to fiddle with PID parameters to find values that work with your application. Simply hook up your mount, set the appropriate limits, and start the AutoTune process. The instrument will automatically calculate PID parameters that will work with your mount.
Simple User Interface
The 6340’s user interface is remarkably simple… so easy to use, you’ll have it up and running in no time. Easy-to-read messages, simple menus, and powerful multi-line display make the instrument incredibly easy to use.
Arroyo Instruments temperature controllers feature bipolar current output, allowing for heating and cooling applications. All Arroyo Instruments temperature controllers, including the OEM packaged 585 and 586 Series TECPak controllers, provide heating and cooling capabilities.
A majority of Arroyo Instruments TEC controllers will support four different temperature sensors: NTC thermistor, RTD, LM335, and AD590 sensors. However, a few TECSource models only support the 10kΩ NTC thermistor.
Yes, Arroyo Instruments TEC controllers offer both low and high temperature set point limits, as well as the ability to fine-tune the controller’s PID parameters. PID parameters can be adjusted using eight standard factory gain settings that are numbered 1 through 300. Increasing the controller’s gain value will increase the speed of your control loop, while a rapid temperature fluctuation can be fixed by reducing the gain value. Users may need to experiment with a few different factory gain values to find the ideal value, and the gain can also be set to PID for even finer tuning. The PID parameters can also be set via the AutoTune Function, where the controller calculates ideal PID values based on your typical operating conditions. Please refer to your temperature controller’s user manual for more information and guidance for starting the Autotune feature.
Yes, most models support USB and RS-232 connections, with a select few models supporting Ethernet. Both the Arroyo benchtop and OEM temperature controllers can be controlled remotely via the ArroyoControl software, or integrated into custom systems using software such as LabVIEW, Python, and MATLAB.
The Arroyo Instruments line of temperature controllers was designed for precision temperature control across a broad range of applications. Common applications and testing environments are laboratory/university research work, laser diode characterization, industrial and manufacturing testing applications, and OEM system integration.
Arroyo Instruments laser drivers support a 50mA to 100A current control range, and up to 56V compliance voltage. If you do not see a current range or compliance voltage that supports your application, please contact [email protected] to inquire about a semi-custom configuration of an existing LaserSource model.
Arroyo Instruments temperature controllers have a temperature stability specification of ±0.004°C over a 1 hour span and ±0.01°C over 24 hours at an ambient room temperature of roughly 25°C.
Specifications
Summary
Laser Driver
4 Amp / 2 Amp, 4 Volt
Temperature Controller
5 Amp, 12 Volt, 60 Watt
Laser Driver
LASER CURRENT
Range (mA)
0 – 4,000 or 0 – 2,000
Resolution (mA)
0.2
Setpoint Accuracy (±[% set + mA])
0.05% + 0.8
Measurement Accuracy (±[% reading + mA])
0.05% + 0.8
Stability (ppm, time)
< 10, 1 hour
Temperature Coeff (ppm/°C)
50
Noise/Ripple (μA rms, low BW)
< 40
Transients (μA)
< 400
Compliance Voltage (V)
4
Modulation Bandwidth (kHz)
150
Modulation Input Range
0 – 10V, 10kΩ
PHOTODIODE CURRENT
Range (μA)
2 – 5,000
Resolution (μA)
0.1
Setpoint Accuracy (±[% set+μA])
0.05% + 1
Measurement Accuracy (±[% reading+μA])
0.05% + 1
Stability (ppm, time)
< 200, 24 hours
Temperature Coeff (ppm/°C)
< 200
PD Bias (V)
0 to -5V, software programmable
LASER VOLTAGE
Range (V)
0 – 4
Resolution (V)
0.001
Setpoint Accuracy (±[% set+V])
0.05% + 0.005
Measurement Accuracy (±[% reading+V])
0.05% + 0.005
Stability (ppm, time)
< 50, 1 hour
Temperature Coeff (ppm/°C)
< 100
Four-wire Measurement
Yes
LIMITS
Current Limit Accuracy (mA)
40
Voltage Limit Accuracy (±% FS)
2.5%
Temperature Controller
Power (A, V, W)
±5A, ±12V, 60W
Stability (1 hour, °C)
0.002
Stability (24 hours, °C)
0.005
Temperature
Range (°C)
-99 to 250
Resolution (°C)
0.01
Thermistor (100µA) Accuracy at 25°C (°C)
0.03
AD590 Accuracy at 25°C (°C)
0.90
LM335 Accuracy at 25°C (°C)
0.90
100Ω RTD Accuracy at 25°C (°C)
0.33
1kΩ RTD Accuracy at 25°C (°C)
0.40
Thermistor, 100µA Range
Accuracy (±[% reading + kΩ])
0.05 + 0.005
Range (kΩ)
0.02 – 45
Resolution (kΩ)
0.001
Thermistor, 10µA Range
Accuracy (±[% reading + kΩ])
0.05 + 0.05
Range (kΩ)
0.2 – 450
Resolution (kΩ)
0.01
LM335
Accuracy (±[% reading + mV])
0.3 + 1
Range (mV)
1730 – 4730
Resolution (mV)
0.1
Bias (mA)
1
AD590
Accuracy (±[% reading + µA])
0.03 + 0.1
Range (μA)
173 – 473
Resolution (μA)
0.01
Bias (V)
4.5
100Ω RTD
Accuracy (±[% reading + Ω])
0.03 + 0.1
Range (Ω)
20 – 192
Resolution (Ω)
0.01
1kΩ RTD
Accuracy (±[% reading + Ω])
0.05 + 1
Range (Ω)
100 – 4500
Resolution (Ω)
0.1
Current
Range (A)
±5
Compliance Voltage (V)
12
Max Power (W)
60
Resolution (A)
0.01
Accuracy (±[% value + A])
0 + 0.03
Noise/Ripple (A, rms)
< 0.005
Current Limit Accuracy (A)
0.05
Voltage (measurement only)
Accuracy (±[% reading + V])
0 + 0.05
Range (V)
±12
Resolution (V)
0.01
General
Laser Connector
DB-9, female
TEC Connector
DB-15, female
Display Type
4×20 VFD
Computer Interface
USB 2.0 Full Speed (USB Type B),RS-232 (DB-9, male)