586-04-56 TECPak, 4A/56V

At-a-Glance.

  • AutoTune automatic PID
  • 4A, 56V (224W) of TEC power
  • 0.004°C stability
  • Bi-polar output: heat & cool
  • Thermistor, LM335, and RTD sensors
  • 4-wire RTD mode

$1,945.00

CONTACT FACTORY FOR LEAD TIME

The 586-04-56 TECPak is a high-performance, compact temperature controller for applications that do not require (or need to omit) a full user interface. Powerful (224W), capable (two sensor inputs, interlock, fan controller), and compact, the 586-04-56 is an excellent mix of features, size, and price.

Excellent Stability

The 586 offers excellent short-term and long-term stability. At less than 0.004°C stability over one hour, and 0.01°C over 24 hours.

Fully Adjustable PID with AutoTune Automatic PID Calculation

Preprogrammed with eight factory set gain settings that support most applications, as well as fully adjustable PID settings. AutoTune for automatic PID calculation simplifies the determination of PID values for your mount.

Simple Multi-Channel System

Using the 1402C-RM Rack Mount Tray, you can quickly rack up to three 586 Paks into a rack for a quick multi-channel system.

586-04-56 TECPak FAQ

Arroyo Instruments temperature controllers feature bipolar current output, allowing for heating and cooling applications. All Arroyo Instruments temperature controllers, including the OEM packaged 585 and 586 Series TECPak controllers, provide heating and cooling capabilities.

A majority of Arroyo Instruments TEC controllers will support four different temperature sensors: NTC thermistor, RTD, LM335, and AD590 sensors. However, a few TECSource models only support the 10kΩ NTC thermistor.

Yes, Arroyo Instruments TEC controllers offer both low and high temperature set point limits, as well as the ability to fine-tune the controller’s PID parameters. PID parameters can be adjusted using eight standard factory gain settings that are numbered 1 through 300. Increasing the controller’s gain value will increase the speed of your control loop, while a rapid temperature fluctuation can be fixed by reducing the gain value. Users may need to experiment with a few different factory gain values to find the ideal value, and the gain can also be set to PID for even finer tuning. The PID parameters can also be set via the AutoTune Function, where the controller calculates ideal PID values based on your typical operating conditions. Please refer to your temperature controller’s user manual for more information and guidance for starting the Autotune feature.

Yes, most models support USB and RS-232 connections, with a select few models supporting Ethernet. Both the Arroyo benchtop and OEM temperature controllers can be controlled remotely via the ArroyoControl software, or integrated into custom systems using software such as LabVIEW, Python, and MATLAB.

The Arroyo Instruments line of temperature controllers was designed for precision temperature control across a broad range of applications. Common applications and testing environments are laboratory/university research work, laser diode characterization, industrial and manufacturing testing applications, and OEM system integration.

Arroyo Instruments temperature controllers have a temperature stability specification of ±0.004°C over a 1 hour span and ±0.01°C over 24 hours at an ambient room temperature of roughly 25°C.

Other Products in this Series

ModelPowerCurrentVoltageStabilitySensors
586-15-28 TECPak, 15A/28V345 W15 A28 V±0.004°CThermistor, LM335, RTD
586-04-56 TECPak, 4A/56V224 W4 A56 V±0.004°CThermistor, LM335, RTD
586-08-26 TECPak, 8A/26V190 W8 A26 V±0.004°CThermistor, LM335, RTD
586-08-54 TECPak, 8A/54V392 W8 A54 V±0.004°CThermistor, LM335, RTD